Calibration Procedure |
A&D |
A & D AD 4321 |
Calibration Procedure for A & D AD 4321 Balance
Turn off Turn throw switch in back press zero CAL1 press stby oper
and you should get ................. in display enter the calibration
calibration weight using the keys place calibration weight on scale
and press stby oper you should get.................. slide switch
back take off calibration weight You be done weight
scales |
A&D |
A&D EK 6000H |
Calibration Procedure for A&D EK 6000H Balance
HOLD SAMPLE AND PRINT CAL OUT CAL 0 PRESS SAMPLE SHOWS Cal press print
cal 0 prees print add calibration calibration weights press print
end remove calibration weight done weight
scales |
A&D |
A&D EP 60 KB |
Calibration Procedure for A&D EP 60 KB Balance
A&D EP 60 KB Throw switch on side cal 0 press print Enter Cal Wt press
tare (possibly twice) calF put on calibration weight press tare Cal
End weight scales |
A&D |
AND FV-30K |
Calibration Procedure for AND FV-30K Balance
OPEN BACK OF SCALE PUSH SWITCH TO "ON" PRESS "UNITS" DISPLAY WILL
READ "CAL O" PRESS "ZERO" PRESS UNITS DISPLAY READS "CAL 3" WAIT FOR
CIRCLE IN UPPER LEFT CORNOR PRESS "ZERO" DISPLAY WILL READ "END" SLIDE
SWITCH BACK weight
scales |
A&D |
AND HV or FW-60K |
Calibration Procedure for AND HV or FW-60K Balance
(FV-30K 60LB/40LB)(FW-10K) CAL 1: MAX CAP. IN KG = 150/60/30KG. CAL
2: 2/3 CAP. IN KG = 100/40/20KG. CAL 3: MAX CAP. IN LB = 300/120/60LBS.
CAL 4: 2/3 CAP. IN LB = 200/80/40LBS. OPEN UP BATTERY COMPARTMENT
REMOVE SCREW FLIP CAL SWITCH TO ON IT WILL READ 9.798 PRESS "MODE/UNITS"
"CAL O" STABLE LIGHT PRESS "REZERO" "CAL1" PRESS "MODE' TO SELECT
SETTING CAL 1,2,3,4 (HIGH/LOW/GO WILL ROTATE SETTINGS) PUT calibration
weight ON (200LBS.) AND WAIT FOR STABLE LIGHT PRESS "REZERO" "CAL
END" IS DISPLAYED PUT SWITCH BACK weight
scales |
Acculab |
ACCULAB S50 AND S100 |
Calibration Procedure for ACCULAB S50 AND S100 Balance
PRESS MODE TO "KG" MODE PRESS AND HOLD "CAL" UNTIL CAL. WT. MODE APPEARS
PRESS "TARE" (WHILE CAL IS STILL FLASHING) "CAL O" WILL BE DISPLAYED
PAUSE 2 SECONDS PLACE CORRECT CAL. calibration weight ON SCALE
"CAL F" DONE WHEN DISPLAY READS CORRECT calibration weight PRESS "TARE"
TO RESET ZERO weight
scales |
Acculab |
Acculab VA8000 |
Calibration Procedure for Acculab VA8000 Balance
Hold F1 & on /off self test -c- press tare f1 cal displayed put on
calibration weight 5000g turn off weight
scales |
Acculab |
Acculab Balance |
Calibration Procedure for Acculab Balance Balance
Empty Scale Press and hold Cal/mode until calibration calibration
weight appears in display Put calibration weight on scale Wait
until a beep. Tthe cal calibration weight will blink and it will go
back into weigh mode weight
scales |
Adam Equipment |
TFW 60 |
Calibration Procedure for TFW 60 Bench Scale
Turn the Power OffTurn the power back on , during the counting from
9 to 0 press func/c key. The display will show "CAL" for a few seconds.
While it is showing "CAL" press the SMPL,PRINT and TARE
keys in sequence to enter the calibration section. The display
will show "unload". Remove any calibration weight from the platform.
Press ZERO/ENTER key. The display will show the last calibration
calibration weight used. If this is correct you cn continue by pressing
the ZERO/ENTER key. If it is not correct use the arrow keys
to change the calibration calibration weight value. When it is
correct press the ZERO/ENTER key. If the calibration is
acceptable the displaywill show "PASS" and then retur to normal. If
an error message is shown try calibration again as a disturbance
may have prevented a successful calibration. weight
scales |
Cardinal |
Cardinal 2235 |
Calibration Procedure for Cardinal 2235 Counting Scale
Remove right side plate& throw switch Press Scale Select for scale
being calibrated Change Calib)? to 1 Press count button till Load___
Load Scale Enter calibration calibration weight Press Count Unload
Scale Press Count Till Done weight
scales |
Denver |
Denver 3000D |
Calibration Procedure for Denver 3000D Balance
hold tare till cal put on 500 or 1000 or 2000 or 3000 in gram calibration
weights cal x cal done weight
scales |
Detecto |
Detecto 2225 |
Calibration Procedure for Detecto 2225 Counting Scale
Turn Scale Off Remove right cover plate Remove blue Jumper in
the corner of the scale attached to the display board (don't confuse
with jumper on main board) Power on scale 7670? press1 then count
cap= (whatever) choose remote or local now if remote just press count
now. If local press scale select then count Press count button
till Load___ Load Scale Enter calibration weight Press Count Unload
Scale Press Count Till Done weight
scales |
Digi |
Digi DI-20 & 28 |
Calibration Procedure for Digi DI-20 & 28 calibration weight
Scale
For the 20 the Jumper JP-2 must be enabled on the main board Press
And hold the rezero button and press the up button to the right of
the tare button then tare twice It then displays the scale increments
change with arrow keys or accept and push rezero Enter the capacity
of the scale push rezero Enter the calibration weight you will
use to cal push rezero cal 0 appears push rezero with nothing on scale
cal sp appears add calibration weight and press rezero then press
tare to get out weight
scales |
Digi |
Digi DC-120,130 ,190 |
Calibration Procedure for Digi DC-120,130 ,190 Counting
Scale
select scale to calibrate using" scale" button Press and hold
the rezero key and press 8715 let go of rezero empty scale press
rezero put calibration weight on scale enter amount of cal wt
press "*" press mode twice do this per scale attached weight
scales |
Digi |
Digi DC-150 |
Calibration Procedure for Digi DC-150 Counting Scale
Press mode repeatedly until maintenance. Press and hold rezero then
press 8715. release rezero press tare (right arrow key) till span
adjust press enter the left of the display is the scale number
to change this press + to the right of that is the scale capacity.
To the riqht of that is the calibration calibration weight. The
calibration calibration weight has to be entered with the same
amount of decimals as the scale capacity. To change either number
use the tare to highiite then enter new ualue. Once all values here
are ok press enter. Empty scale push rezero add calibration weight
push enter * press mode till operation press enter weight
scales |
Digi |
Diqi DC-60 |
Calibration Procedure for Diqi DC-60 Counting Scale
Remove label casette In the back of opening there is a 3 position
switch. Move this switch into the middle position. Select scale
to calibrate using scale button. Turn key to program Press and
hold rezero while pressing 8715 release rezero. Empty scale push
rezero add calibration weight enter calibration weight push enter
turn key back to operate move switch back to original position
weight scales |
Doran |
Doran 8000 |
Calibration Procedure for Doran 8000 Bench Scale
Doran 8000 Remove back cover plate Throw red switch inside Displays
AUGAI Press "conv" Display ENT 0 Put calibration weight on scale
zero=up tare = down to correct calibration weight Press "conv" displays
ent sp Take calibration weight off flip switch back weight
scales |
Doran |
Doran 8000XL |
Calibration Procedure for Doran 8000XL Bench Scale
Doran 8000 xl press cal sw if cap ok press units cal 0 press zero
put calibration weight on scale press units press print hold print
to go lower or capicity to go higher till cal calibration weight is
matched press zero done |
Doran |
4200 |
Calibration Procedure for 4200 Bench Scale
remove small plate in back of scale. thow red switch press gross
net c000.00 press zero press units add calibration weight use arrows
to correct units to save gross net weight
scales |
DORAN 7000 |
DORAN |
Calibration Procedure for DORAN Bench Scale
OPEN UP SQUARE IN BACK (RECTANGULAR) "FLIP RED SWITCH" "CO" PRESS
"REZERO" (COUNTS DOWN) C=calibration weight VALUE PUT THAT AMOUNT
ON PRESS "REZERO" (COUNTS DOWN) C=calibration weight VALUE PUT THAT
calibration weight ON PRESS "REZERO" (COUNTS DOWN) DONE FLIP SWITCH
BACK weight scales |
Fairbanks |
Fairbanks HR-2300-2 |
Calibration Procedure for Fairbanks HR-2300-2 calibration
weight Scale
Fairbanks HR-2300-2 Press Program press ^ till cal code is 113 then
enter wait for scale I forgot from there weight
scales |
Fairbanks |
Fairbanks UPS |
Calibration Procedure for Fairbanks UPS calibration weight
Scale
Fairbanks ups scale remove plate on left side press button press
zero press units place displayed calibration weight on scale press
cal button weight scales |
Fairbanks |
Eclipse PPI-HR5000-X |
Calibration Procedure for Eclipse PPI-HR5000-X Bench Scale
remove back cover press black button on board twice to get to CC CAL
Hit units when reads 0 add calibration weight. Hit units again. Adjust
to correct calibration weight using zero/AT sw on side of unit then
hit units again weight
scales |
FIling |
Filing DR-2100A |
Calibration Procedure for Filing DR-2100A Fork Truck Scale
Filing DR2100A 9090 ENTER DEL TOGGLES Y N MENU ENTER FOLLOW INST IF
CAL SW OFF OPEN BOX AND THROW SWITCH weight
scales |
Filing |
Filing DR-5000 |
Calibration Procedure for Filing DR-5000 Fork Truck Scale
Filing DR-5000 The F1 & F2 buttons are shut off points for filling
applications Cal Procedure To get into calibration mode press
"9","9","Re-Zero" This calls the true zero. Go to the pot labeled
R-40 and get as close as possibleto zero Push zero button Apply load
Go to R-41 and dial to correct calibration weight Remove calibration
weight Press "Re-zero" Re adjust if needed Once that is accomplished
Call up true zero again. Go to R-40 and turn to negative 1/2 the scale
capacity (-2500 to -2600lb) then rezero with button re apply calibration
weight if incorrect, turn fine span pot 1/4 clockwise for up or 1/4
counter clock wise for down Do not watch the indcator because this
will screw you up. Re zero the scale and re apply the calibration
weight. repeat the procedure until it is correct. weight
scales |
Flexweigh |
DWM IV |
Calibration Procedure for DWM IV Indicator
Throw SW1 ->FSE F1 Push 5 FSE F6 Press Print F6.0 0 Print Print Print
f6.3d Tareprint F6.40 print Escale print capacity shows, press enter
add load after countdown add load push print enter calibration weight
push enter weight scales |
GSC Scale |
GSC Scale |
Calibration Procedure for GSC Scale Indicator
Turn On, press ¨Stanby¨, after press¨ZERO¨hold and press ¨STANBY¨Display
will ready ¨SYS¨. Press ¨ZERO¨, display will ready 00000, the las
digit will be blinked, Press ¨P¨ to increase to 6, press ¨NET/GROSS¨
to move to the left, press ¨P¨to increase to 5, press ¨NET/GROSS¨
to move to the left, press ¨P¨ to increase to 4, the display will
read 00456, Press ¨ZERO¨, the display will show F1, press ¨SET¨,the
display will show ¨F0¨, press ¨SET¨, the display will show CAL, press
¨ZERO¨, remove load from the balance, press ¨NET¨, Press ¨P¨ to include
the value of weigths that will be use for calibration. Put Load
on the scale and press¨ZERO¨, wait just a moment. DONE. weight
scales |
GSE |
GSE-350 |
Calibration Procedure for GSE-350 calibration weight Scale
GSE 350 Indicator Press zero and select simultaneously press zero,print,units
and tare one after the other scale should say first zero. Press
tare with nothing on the scale use th print and units keys to
enter the load calibration weight press tare Add calibration weight
press tare press tare till out weight
scales |
GSE |
GSE 450 460 |
Calibration Procedure for GSE 450 460 calibration weight
Scale
GSE 450 & 460 Indicator 1) Press [Zero]+[Select] Simultaneously 2)
Press [Zero] [PRINT] [UNITS] [ENTER] In that order 3) Remove any extraneous
Load 4) Press[Select] until "New Zero" is dislayed 5) Press[Enter]
6) place cal calibration weight on scale 7) Key in the calibration
weight value 8) Press [Enter] "Cal OK" 9) Press [Enter] 3 times
weight scales |
GSE |
GSE 455 465 550 |
Calibration Procedure for GSE 455 465 550 calibration weight
Scale
GSE 455-465-550-574-650-660 1) From the calibration weight screen
keyy in 100 [Select] DIsplays Key in Code: The access code must be
keyed in within 5 seconds 2) Press 54321 [ID] [Enter] 3) Press Enter
4) Apply cal calibration weight 5) Key in calibration weight and press
enter 6) Cal OK Press [Enter] 7) Enter=Save press [Enter] 8) Enter=Exit
[Enter] weight scales |
GSE |
GSE 625 622 |
Calibration Procedure for GSE 625 622 calibration weight
Scale
GSE 625 or 622 Counting Scale Use the scale select key to
choose the scale you wish to calibrate Press the [GSE][MODE] to
put the instrument into offset display mode press the [GSE][ZERO].
The scale is now in the auto cal mode Before going any further
enter the program jumper bypass code [23640][GSE][.] The lower display
will prompt you through the set up of Auto Cal. First of all remove
all calibration weight and press [Zero]. The scale will average
the zero while it counts to 10 in the lower display. Counter will
reseet when motion is sensed When tho count reaches 10 place the calibration
weight on scale,enter that amount and press[ENTER] When the procedure
is complete press[MODE] to return to offset displayed If you want
to cal another scale use scale select and start all over or
press [ZERO] to re-enter the Weigh Mode weight
scales |
Masstron |
Masstron M -1500 |
Calibration Procedure for Masstron M -1500 calibration
weight Scale
Masstron M1500 Open Indicator and there is a fine span pot labeled
R-17 Little black pot in middle of board Medium span SW3 M5000 Open
indicator and main group of boards is on a hinge. undo screw in back
so boards swing out. Then throw SW!-1 on the back board. then on keypad
type "access" "0" "5"then press enter this is the zero cal. astrange
number is there press enter till done. For span press "access" "0"
"4" then enter. place calibration weights on scale. tell the indicator
what it is then press enter. To save changes press access then enter
twice then wait two seconds before throwing SW1-1 back. weight
scales |
NCI |
NCI Counting Scale |
Calibration Procedure for NCI Counting Scale Counting
Scale
NCI Counting scale Go to scale to calibrate Take off cover
remove small cover plate throw sw2 on s2 to closed put on 1 calibration
weight enter value and press acc+ Do 2nd value the same way then throw
back switch weight
scales |
Ohaus |
Ohaus DS-10 |
Calibration Procedure for Ohaus DS-10 Balance
Ohaus DS-10 110 LB Capacity Hold button for 15-20 sec 0 Cal in window
place 110 LB on scale it will say 50 KG Remove calibration weight
weight scales |
Pennsylvania |
Pennsylvania 2720 |
Calibration Procedure for Pennsylvania 2720 Counting Scale
PENNSYLVANIA 2720 20 LB CAPACITY SCALE: PRESS "2" PRESS "ENTER"
ADD .5 THEN PUT ON 10 LBS. PRESS "ENTER" ADD 1 THEN PUT ON 20 LBS.
PRESS "ENTER" ............GOES BLANK THEN TAKE calibration weight
OFF PRESS (KEY IN) "20" ALL ZEROES INCLUDED PRESS "ENTER" PRESS ".005"
PRESS "ENTER" (*SAME BELOW FOR SCALE 2) PRESS "2.20462" PRESS
"ENTER" TURN KEY BACK SCALE 2 100 LB. CAPACITY: PRESS "2" PRESS
"ENTER" ADD .5 PUT ON 50 LBS THEN "ENTER" ADD 1 PUT ON 100LBS. THEN
"ENTER" .......... TAKE calibration weight OFF PRESS "100.00" THEN
"ENTER" GO BACK TO * weight
scales |
Pennsylvania |
Pennsylvania 420-10 |
Calibration Procedure for Pennsylvania 420-10 Counting
Scale
PENNSYLVANIA 420-10 (10/200 CAP) SAMPLE=SMALL PLATFORM BULK=LARGE
PLATFORM TURN KEY TO PROGRAM KEY IN SENSITIVITY OF LOAD CELL "2" THEN
"ENTER" ADD .5 (1/2 CAP. OR KEY IN OTHER calibration weight) PUT calibration
weight ON PRESS "ENTER" ADD 1 (FULL CAPACITY OR ENTER IN AMOUNT) PUT
calibration weight ON PRESS "ENTER" .......WILL STROBE AND THEN BLANK
OUT TAKE calibration weight OFF .......WILL STROBE AGAIN AND THEN
BLANK OUT (POT SCREW TO 1.5 MV OR LIFT UP CORNOR OF PLATFORM) KEY
IN CAPACITY OF LOAD CELL INCLUDING ALL ZEROS (2OO.OO) PRESS "ENTER"
BLANK OUT .05 RESOLUTION KEY IN PRESS "ENTER" ONLY FOR SAMPLE SCALE:
PRESS "2.20462" PRESS "ENTER" weight
scales |
Pennsylvania |
Pennsylvania 5400 |
Calibration Procedure for Pennsylvania 5400 Counting Scale
Pennsylvania 5400 Counting 1)Turn Key in Back of scale 2) Enter
scale number to be calibrated 3) Press enter till cal 18 STOP
4) Wait about 10 sec 5) Press Enter 6) Press enter again 7) Empty
scale and press enter 8) Put 1/2 scale capacity on scale
and press enter OR put different calibration weight on scale and
enter value then press enter 9) Put Full scale capacity calibration
weight on scale or enter value and press enter 10) Press eter
till STOP 11) Turn Key Bac kweight
scales |
Pennsylvania |
Pennsylvania 5600 |
Calibration Procedure for Pennsylvania 5600 calibration
weight Scale
Penn 5600 hold down 4 key full capicity press enter till cal type
use units to change to load cell press enter follow instructions or
enter different calibration weight values followed by enter weight
scales |
Pennsylvania |
Pennsylvania 7500 |
Calibration Procedure for Pennsylvania 7500 Bench Scale
Penn 7500 push button under plate under platter. The plate is held
in place by a screw on the bottom of the scale. Push the button
under the cover. Press print till cal 42 press print to cal zero.
Put on half capacity and press print. Cal 44 put on full capacity
and press print. Cal zero print . then print to cal 99 and the scale
will count down and all done. weight
scales |
Rice Lake |
Rice Lake CW-40 |
Calibration Procedure for Rice Lake CW-40 Bench Scale
Rice Lake CW-40 Flip Red Switch behind access plate Press under button
Display C press zero for zero cal scale says ent zero add calibration
weight and adjust with up and down arrows. Press Units Flip switch
back weight scales |
Rice Lake |
Rice Lake CW-80 |
Calibration procedure for Rice Lake CW-80 Bench Scale
Press config Switch
Right over to calibration
S croll down to zero
C lear the platter and press the zero
witll display zcal for 10 sec
when done calsp
display shifts to full scale cap weight
adjust weight and press zero
scal is displayed while working
CalIB means it is done
remove weights
Press config switch
you are done
|
Rice Lake |
Rice Lake IQ-600HB |
Calibration Procedure for Rice Lake IQ-600HB Weigh Scale
Rice Lake IQ 600 HB TAKE DOWN METAL BAR IN FRONT 4 SWITCHES FLIP SWITCH
3 1234 (GET READING AS CLOSE TO ZERO W/PENTIAMETER) FLIP SWITCH BACK
PRESS "ZERO" CLOSE SWITCH #1 (MAY NOT BE ZERO THAT`S OK) PRESS "ZERO"
WITH NO calibration weight ON "CAL O" "C" PLACE 10% ON SCALE PRESS
(KEY IN) 500 LBS. (NET/GROSS =UP TARE=DOWN) PRESS "CONV" FLIP SWITCH
BACK weight scales |
Rice Lake |
IQ 6500 6200 9500 |
Calibration Procedure for IQ 6500 6200 9500 Counting Scale
Rice ake IQ 6200, 6500, 9500 Press and hold the rezero key and press
8715 let go of rezero press rezero put calibration weight on scale
enter amount of cal wt press "*" press mode twice do this per scale
attached |
Rice Lake |
Rice Lake RS-130 |
Calibration Procedure for Rice Lake RS-130 click
here |
Sartorius |
CP3000 |
Calibration Procedure for CP3000 Balance
Hold Down Tare Apply calibration weight Displayed Follow Instructions
weight scales |
Scientech |
Scientech |
Calibration Procedure for Scientech Bench Scale
Scientech Balance 1) Press the FUNCTION switch (or MDE and RANGE switches
simultaneously for dual range models) 2) Press the MODE switch when
CAL1 appears. 3) Wait 10 seconds. Then press the ZERO switch 4) Place
one of the flashing calibration calibration weights on the pan
5) ait 10 seconds. Then press the zero switch 6) Remove calibration
weight weight scales |
Setra |
Setra |
Calibration Procedure for Setra Counting Scale
Setra counting scales Press "-" press"clear" press "zero" put calibration
weight on scale enter calibration weight value press "set up/cal"
weight scales |
Sterling |
Sterling |
Calibration Procedure for Sterling Counting Scale
Sterling 825 Counting Scale 1)Press Scale Select to go to
the scale you are calibrating 2) Press and hold Gross/Net Key
until Params 3) Now the NTEP parameter along with its value is displayed
4)Press the print key or zero key till FCAL 5)Select FCALT with tare
key and press Print 6) Shows current number f scales to change tis
pram press tare or press print to move on 7) This is capacity change
it or press print to move on 8)Graduations press print or change 9)Shows
Cal Units print to move on 10) CAL F Change this to CAL T using Tare
button press print to begin cal a)Clear scale and press print
b)Step A8 press print c)Place calibration weight on scale, enter
calibration weight value and press print 11)SPN 0.0 this is the span
error press print 12)Shows caacity of next scale. Repeat steps
7-11 13)Gross/Net or reset key will get you out at the end weight
scales |
Toledo |
Toledo 8571 |
Calibration Procedure for Toledo 8571 Counting Scale
TOLEDO 8571 PRESS "ENTER" UNTIL E SCALE (F 3.2 F 3.3) AFTER E
SCALE PRESS "*" (PRINT) ADD LD (TARE=CHANGE #'S CLEAR=NEXT # ZERO=BACKS
UP) "OOO" ADD calibration weight PRESS "*" (PRINT) PRESS "CLEAR" PRESS
"*PRINT" UNTIL F 4.5 THEN "*" (PRINT) SET OFF weight
scales |
Toledo |
Toledo 8581 |
Calibration Procedure for Toledo 8581 calibration weight
Scale
Toledo 8581 Counting Scale Throw Switch under platter Press 0
to F3 Press print to F3.2 Press1 Scale Cap Displayed Press Print
Increment Size Press Print E Scl (Empty Scale) Press Print Scale
Counts Down Add calibration weight Press Print Enter calibration weight
in whole numbers Press Print Counts Down Cal d (Done) Press print
till set up off Throw switch back Turn scale off & on weight
scales |
Toledo |
Toledo SC-05 |
Calibration Procedure for Toledo SC-05 calibration weight
Scale
Toledo SC-05 lift platter. Press cal switch in hole to the right of
platter. Puts yuu into service. Use rezero enter and tare to get into
cal mode weight scales |
Transcell |
TI-500E |
Calibration Procedure for TI-500E Scale Indicator
Turn Off Scale. Throw Switch on back of scale. Turn On Scale.
F1 press right arrow till F16. Press Down Arrow. C0 Press Zero Key.
Press Net/Gross. Says Set. F17 with right arrows. Arrow down. C1 then
flash 0. Add calibration weight. Change weigh using arrow buttons.
Press Gross/Net end C1 Turn off and throw switch back weight
scales |
Vibra Screw |
Filling System |
Calibration Procedure for Filling System Filling System
Throw cal switch on dip switches located at the bottom of aluminum
box. Press mode to get to gross mode press the 8 button (also says
cal). Displays SPCd press enter enter 5321 then enter setup press
enter till cal press enter Hd Cal Press enter "Zero" displays press
enter "-" displays press test clr 0.0 displays press enter span displays
place calibration weight on scale press enter "-" displays press
test/clr enter calibration weight press enter lincor press enter 0.0
displays press enter return displays press enter end cal displays
press enter switch back weight
scales |